<link rel="stylesheet" href="styles.f3b1fba60ec7970c.css">

Desenvolvimento de hardware para análise de bioimpedância em banda larga

Carregando...
Imagem de Miniatura

Título da Revista

ISSN da Revista

Título de Volume

Editor

Universidade Federal do Rio de Janeiro

DOI

Resumo

The bioimpedance spectroscopy is a technique that allows the biological tissue research from its electrical properties. The applications include body composition, research in cancer cells, bioimpedance tomography and others. This paper realizes the development of a bioimpedance analyzer for frequencies between 1 kHz and 1 MHz and impedance range between 50 and 1 kΩ. The system perform the acquisition with a tetrapolar topology, which a current excitation used was ternary chirp signal. For device control and visualization of the results, a software was also developed in LabVIEW platform. As a result, the project requires a total time of 2.47 s to perform the bioimpedance measurement, however only 10 ms are used for tissue excitation, the resolution obtained was 2.44 Ω for the module and 0.29° for the phase, with errors of accuracy around 30 Ω (module) and 6 ° (phase).

Descrição

Citação

Avaliação

Revisão

Suplementado Por

Referenciado Por

Direitos e licensiamento