Desenvolvimento de hardware para análise de bioimpedância em banda larga
Carregando...
Arquivos
Data
Título da Revista
ISSN da Revista
Título de Volume
Editor
Universidade Federal do Rio de Janeiro
DOI
Resumo
The bioimpedance spectroscopy is a technique that allows the biological tissue
research from its electrical properties. The applications include body composition,
research in cancer cells, bioimpedance tomography and others.
This paper realizes the development of a bioimpedance analyzer for frequencies
between 1 kHz and 1 MHz and impedance range between 50 and 1 kΩ. The system
perform the acquisition with a tetrapolar topology, which a current excitation used was
ternary chirp signal. For device control and visualization of the results, a software was
also developed in LabVIEW platform. As a result, the project requires a total time of
2.47 s to perform the bioimpedance measurement, however only 10 ms are used for
tissue excitation, the resolution obtained was 2.44 Ω for the module and 0.29° for the
phase, with errors of accuracy around 30 Ω (module) and 6 ° (phase).